ENGLISH
!
The Sleepnet Corporation Mojo
such as CPAP or bi-level, operating at or above 3 cm H2O for the treatment of adults.
CAUTION: U.S. Federal Law restricts this device for sale by or on the order of a physician.
CAUTION: Do not expose the Mojo
CAUTION: This mask is intended for single patient use only. It may be cleaned and used repeatedly on the
same person, but must not be used on multiple persons. Do not sterilize or disinfect.
WARNING: Do not block or try to seal the vent holes or anti-asphyxia valve.
WARNING: At low CPAP pressures, the fl ow through the exhalation ports may be inadequate to clear all
exhaled gas from the tubing. Some rebreathing may occur.
WARNING: Do not use the Mojo
WARNING: This mask should be used with the positive airway pressure device recommended by the manu-
facturer or by your physician or respiratory therapist. The mask should not be used unless the positive airway
pressure device is turned on and operating properly. The mask's vent holes should never be blocked. The vent
holes allow a continuous fl ow of air out of the mask. When the positive airway pressure device is turned on and
functioning properly, new air from the device fl ushes the exhaled air out through the mask's vent holes. When the
positive airway pressure device is not operating, exhaled air may be rebreathed. Rebreathing of exhaled air can,
in some circumstances, lead to suffocation. This warning applies to most models of CPAP machines and masks.
WARNING: If oxygen is used with this CPAP mask, the oxygen fl ow must be turned off when the CPAP
machine is not operating. Explanation of the warning: When the CPAP device is not in operation, and the oxygen
fl ow is left on, oxygen delivered into the ventilator tubing may accumulate within the CPAP machine enclosure.
Oxygen accumulated in the CPAP machine enclosure will create a risk of fi re.
WARNING: No smoking or open fl ames such as candles may be used when oxygen is in use.
WARNING: Consult manufacturer for proper O2 enrichment tubing. At a fi xed rate of supplemental oxygen
fl ow, the inhaled oxygen concentration will vary, depending on the pressure settings, patient breathing pattern,
mask selection, and the leak rate.
WARNING: To minimize the risk of vomiting during sleep, avoid eating or drinking for three (3) hours before
using the mask.
WARNING: Any unusual skin irritation, chest discomfort, shortness of breath, gastric distension, abdominal
pain, belching or fl atulence from ingested air, or severe headache experienced by the patient during or immedi-
ately after use should be immediately reported to a healthcare professional.
WARNING: Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition. If
symptoms occur, consult a physician or dentist.
WARNING: Consult a physician if any of the following symptoms occur while using the mask: Drying of the
eyes, eye pain, eye infections, or blurred vision. Consult an ophthalmologist if symptoms persist.
WARNING: The elbow and anti-asphyxia valve assembly have specifi c safety functions. The mask should not
be used if the valve fl apper is damaged, distorted, or torn.
WARNING: This mask is not intended for patients without a spontaneous respiratory drive. This mask should
not be used on patients who are uncooperative, obtunded, unresponsive, or unable to remove the mask. This
mask may not be suitable for persons with the following conditions: impaired cardiac sphincter function, exces-
sive refl ux, impaired cough refl ex, and hiatal hernia.
WARNING: Do not use in an MRI environment.
WARNING: Magnets are used in the mask and headgear clips with a fi eld strength of 380mT. With the excep-
tion of the devices identifi ed in the contraindication, ensure that the mask is kept at least 6 inches (approx.16
cm) away from any other medical implants or medical devices that can be impacted by the magnetic fi elds to
avoid possible effects from localized magnetic fi elds. For instance, the functioning of implants may change, or
implants may move within the body due to magnetic attraction/repulsion. This applies to you or anyone in close
physical contact with your mask.
• Do not use this product if you are experiencing nausea, vomiting, taking a prescription drug that may cause
vomiting, or if you are unable to remove the mask on your own.
• Do not use this mask if you or anyone (example: household members, bed partners, caregivers, etc.) in close
8
Mojo
Vented Full Face Mask
®
CONTENTS: Mojo
®
Rx ONLY
DOES NOT CONTAIN NATURAL RUBBER LATEX
®
Full Face Mask is intended to be used with positive airway pressure devices,
GENERAL INFORMATION - WARNINGS AND CAUTIONS
Full Face Mask to temperatures above 122°F (50°C).
®
Full Face Mask with baseline pressure less than 3 cm H2O.
®
CONTRAINDICATIONS
Vented Full Face Mask with Headgear
MADE IN U.S.A.