Notes
1) In the USA: The installation must be in accordance with the National Electrical Code
505 and ANSI / ISA-RP 12.6.
In Canada: The installation must be in accordance with the Canadian Electrical Code
2) The apparatus must not be connected to any device that uses or generates in excess of 250Vrms or DC.
U
= 250V.
m
3) In the USA: The Apparatus must be connected to a suitable ground electrode per National Electrical Code
NFPA 70, Article 504 or 505. The resistance of the ground pad must be less than 1 ohm.
In Canada: The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code
Part 1. The resistance of the ground pad must be less than 1 ohm.
4) Connection by non interchangeable cable type LiYC-Y-CY 4 x 0.5; max length: 50m (164 ft).
5) Connection by means of polarized connector outside of the indicator.
6) Connection by non interchangeable cable type LiYC-Y-CY 4 x 0.5; max length: 0.5m (1.6 ft).
8) The circuits of the data output circuits shall be assumed to be connected to earth.
9) The cable needs not be protected against damage.
10) Equipment with metallic housing (IP4X in minimum) and passive wiring, only. For use in Class II,III, Division 1
and in Zone 21 the housing must be IP6X.
11) The cable to the T-Connector must be protected against damage.
12) Any NRTL approved equipment with Entity Concept parameters (see note 13)
13) The Entity Concept allows interconnection of intrinsically safe apparatus with associated apparatus not
specifically examined in combination as a system when the approved values of Voc, Isc and Pmax resp. Uo, Io,
Po of the associated apparatus are less than or equal to Vmax, Imax and Pmax resp. Ui, Ii, Pi of the intrinsically
safe apparatus and the approved values of Ca and La resp. Co and Lo of the associated apparatus are greater
than Ci and Li of the intrinsically safe apparatus plus all cable parameters.
For the input and output parameters of the data interface of the CAIXS2 see page 2.
14) The weighing unit series SIWXS...-3-..-... and SIWXS...-3-V... (complete scales) and ISX...-..-... and ISX...-V...
(digital weighing platforms) are approved/certified by FM for use in the USA and in Canada. See Certificate of
Compliance and Control Drawing number 36953-751-07-A4.
15) The Interface Converter YDI05-Z.. is approved/certified by FM for use in the USA and in Canada. See Certificate
of Compliance and Control Drawing number 65710-800-07-A4.
16) The Power Supply Model YPS02-ZKR is approved/certified by FM for use in the USA and in Canada. See
Certificate of Compliance and Control Drawing number 65501-000-17.
17) The rechargeable battery pack YRB02-X is approved/certified by FM for use in the USA and in Canada. See
Certificate of Compliance and Control Drawing number 65656-000-07-A4.
18) The Power Supply Model YPS02-XKR is /certified by CSA for use in Canada. See Certificate of Compliance and
Control Drawing number 65516-000-17.
19) The Power Supply Model YPS02-XUR is approved by FM for use in the USA. See Certificate of Compliance and
Control Drawing number 65516-000-17.
20) The output parameters of the Ex ia circuit to the load cells are:
Uo = 11.8 V, Io = 147 mA, Po = 1.49 W, Co = 770 nF, Lo = 300 µH
21) See entity parameters on the control drawing 35739-003-07-A4 or 65607-000-07-A4 of the weighing platform
types CAPX..U-......-......, IUX.4-......-..... and IFX.4-......-.....
22) Ambient temperature range: -10°C .... +40°C (14°F .... +104°F) The temperature class for gases of the CAIXS2
is T4. The maximum surface temperature for dusts of the CAIXS2 is 80°C (176°F).
23) WARNING:
AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA SÉCURITÉ
2018-08-27
T. Hiller
176
Mode d'emploi Combics CAIXS2
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
INTRINSÈQUE.
®
, NFPA 70, Article 504 or
®
, Part1, Section 18.