ADVENTURER BALANCES
9.3 Communication
9.3.1
Interface Commands
Commands listed in the following table will be acknowledged by the balance.
The balance will return "ES" for invalid commands.
Command
Characters
IP
P
CP
SP
SLP
SLZP
xP
0P
H
Z
T
xT
PT
ON
OFF
C
IC
AC
PSN
PV
x#
P#
x%
P%
xS
xRL
PFMT [n];
PFMT space N
9.3.2
RS232 (DB9) Pin Connections
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
9.4 The USB Interface
The Ohaus USB Interface is a unique solution to the problem of connecting a balance to a computer using a Universal
Serial Bus (USB). USB devices are categorized into classes such as disk drives, digital cameras, printers, etc.
Balances do not have a commonly used class so the Ohaus USB interface uses a generic interface based on the
RS232 serial standard.
Data sent from the balance to a computer is in USB format. The USB data is directed to a virtual port. This port then
appears as an RS232 port to the application program.
Find Quality Products Online at:
TABLE 9-6. ADVENTURER INTERFACE COMMAND LIST
Function
Immediate Print of displayed weight (stable or unstable).
Print displayed weight (stable or unstable).
Continuous Print.
Note: for AX...N...type models, when LFT is turned ON, CP could not work.
Print on Stability.
Auto Print stable non-zero displayed weight.
Auto Print stable non-zero weight and stable zero reading.
Interval Print x = Print Interval (1-3600 sec) 0P ends interval Print
See above
Enter Print Header Lines
Same as pressing Zero Key
Same as pressing Tare Key.
Establish a preset Tare value in displayed unit. X = preset tare value. Sending 0T clears
tare (if allowed).
Prints Tare weight stored in memory.
Brings out of Standby
Goes to Standby.
Begin Span Calibration
Begin internal Calibration, same as trigger from calibration menu.
Abort Calibration. Attention: when LFT ON, the operation is not allowed.
Print Serial Number.
Print terminal software version, base software version and LFT ON (if LFT is set ON).
Set Counting APW (x) in grams. (must have APW stored)
Print Counting application APW.
Set Percent application reference weight (x) in grams. (must have reference weight stored)
Print Percent application reference weight.
0 = print unstable data, same as IP; 1 = print stable only
disable response;
enable response. This command only controls the "OK!"
response.
n=0 default OHAUS mode
n=1 MT mode
n=2 Sartorius mode
GlobalTestSupply
www.
, same as SP.
1)
Data
GND
1
5
9
6
Handshake
.com
sales@GlobalTestSupply.com
EN-61