EN-36
5.4 Weighing Units
Enter this sub-menu to activate the desired units of measure. This menu can also be accessed by
pressing the unit symbol in an application home screen.
Note: Due to national laws, the balance may not include some of the units of measure listed.
Find Quality Products Online at:
5.3.9
Approved Mode
Use this menu to set the Legal for Trade status.
OFF = standard operation.
ON = operation complies with Legal Metrology regulations.
Note: When Approved Mode is set to ON, the menu settings are affected as follows:
Calibration Menu:
Automatic Calibration is forced to be turned ON and hidden. Internal Calibration
and Calibration Test functions are available. All other functions are hidden.
For AX...N... models:
Automatic Calibration will be locked at its current setting.
-
If you set Internal Calibration to be On before you turn on Approved Mode,
-
Internal Calibration menu will still be available.
If you set Internal Calibration to be Off before you turn on Approved Mode,
-
Internal Calibration menu will be locked.
Balance Setup Menu:
Filter Level is locked at the current setting.
Auto Zero Tracking is limited to 0.5 Division and OFF. The selected setting is
locked.
Auto Tare is locked at current setting.
Graduations is forced to 1 Division and the menu item is hidden.
For AX...N... models, graduations will be locked at its current setting.
Communication Menu (Communication->Print Settings->Print Output):
Stable Weight Only is locked ON.
Numeric Value Only is locked OFF.
Communication Menu (Communication->Print Settings->Auto Print):
For AX...N...models, auto print mode selections are limited to OFF, On Stability,
and Interval. Continuous is not available.
Data Maintenance Menu:
Export to USB is hidden
Import from USB is hidden
Lockout Menu:
Menu is hidden
Note: The security switch located at the rear of the balance must be in the locked
position to set Approved Mode to ON. The security switch must be in the unlocked
position to set Approved Mode to OFF. See Section 6.
Weighing application main screen with LFT turned ON.