Télécharger Imprimer la page

CooperSurgical Endosee ES9000 Manuel De L'utilisateur page 38

Publicité

All manuals and user guides at all-guides.com
Endosee
System - User Manual
®
Professional Healthcare Facility Environment
Immunity Test
IEC60601 Test
and Compliance
Electrical Fast
+/- 2 kV for ac
Transient/Burst
power supply
lines
IEC 61000-4-4
+/- 1 kV for input/
output lines
100 kHz repetition
frequency
Surge
+/- 0.5, 1 kV line(s)
to line(s)
+/- 0.5, 1, 2 kV
line(s) to earth
IEC 61000-4-5
Conducted
Professional
disturbances
Health Care
induced by RF
3V
fields
0.15-80 MHz
6V in ISM bands
IEC 61000-4-6
between 0.15
MHz and 80 MHz
80% AM at 1kHz
Voltage dips
0 % U
0.5 cycle
IEC 61000-4-11
0, 45, 90, 135,
180, 225, 270
and 315°
0 % U
1 cycle 0°
70 % U
25/30 cycles 0°
Voltage
0 % U
Interruptions
300 cycles 0°
IEC 61000-4-11
Power frequency
30 A/m
(50/60 Hz)
50 or 60 Hz
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic
environment –
guidance
for
T
for
T
for
T
for 250/
T
38

Publicité