Endosee
System - User Manual
®
5.4 Sterile, Single-use Disposable Cannula
The disposable Cannula contains a miniature CMOS
camera and light-emitting diode (LED) illumination mod-
ule at the tip. The cannula connects
to the Handheld Display Module through an electrical
connector for: DC power, image data transfer and
removal and control of the CMOS camera and
illumination module. The cannula has an inner channel
for the infusion of fluid. Fluid infusion is achieved by
attaching either a pre-filled syringe or a pressurized bag
of distending medium to the inflow port through
an extension tube or a tubing set. The outflow port is at
the distal tip of the cannula and empties out around the
tip to infuse fluid into the uterine cavity or female bladder.
The cannula has a pre-curved tip section at an angle of
20°± 5°. The cannula tip section should not be altered.
The cannula maximum temperature will not exceed 41°C.
Figure 6 shows the cannula.
3
4
1.
Handheld Display Module Connector
2.
Inflow Port
3.
Tip Section
4.
CMOS Camera & LED Illumination Module
Figure 6: Disposable Cannula
1
2
19