Endosee
System - User Manual
®
8.2 Guidance and Manufacturer's Declaration –
Electromagnetic immunity
The Endosee System is intended for use in the electromagnetic
environment specified below. The customer or the user of
the Endosee System should assure that it is used in such
an environment.
Professional Healthcare Facility Environment
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Radiated RF
EM Fields
IEC 61000-4-3
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
IEC60601 Test
Electromagnetic
and Compliance
environment –
guidance
+/- 8 kV contact
+/- 15 kV air
Professional
Health Care 3V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
9 V/m
WARNING: Portable
710,745,780 MHz
RF communications
Pulse modulation
equipment (including
217 Hz
peripherals such as
antenna cables and
9 V/m
external antennas)
5240,5500,
should be used no
5785 MHz
closer than 30 cm (12
Pulse modulation
inches) to any part of
217 Hz
the Endosee
System, including
27 V/m
cables specified by
385 MHz
the manufacturer.
Pulse modulation
18Hz
28 V/m
450 MHz
FM +/- 5 kHz
Otherwise,
deviation; 1 kHz
degradation of the
sine
performance of this
equipment could
28 V/m
result.
810, 870, 930 MHz
Pulse modulation
Interference may
18 Hz
occur in the vicinity of
equipment marked
28 V/m
with the following
1720,1845,1970,
symbol:
2450 MHz
Pulse modulation
217 Hz
37
®