1. Long press the Rotation knob to enter the menu.
2. Navigate to
3. Remove pan holder and press knob.
4. Place temperature kit and press knob.
5. Close cover and press knob.
Note:
In this first session, the dryer will heat and stabilize at 100°C for 15 minutes.
6. When the first session ends, the screen will display "Enter temp reading from kit.
Check the temperature measured by the Mechanical Temperature Calibration Kit, then
rotate the knob to enter the temperature reading.
7. Press the knob to confirm. The second adjustment session will start.
Note:
In the second session, the dryer will heat and stabilize at 160°C for 15 minutes.
8. When the second session ends, the screen will display "Enter temp reading from kit.
Check the temperature measured by the Mechanical Temperature Calibration Kit, then
rotate the knob to enter the temperature reading.
9. After enter the temperature reading, the screen will display "Adjustment done,press
tare.
Press
to return to the Home screen.
5.3.2.2. View Temperature Adjustment History
MB62 stores the latest valid temperature history.
To navigate to the Temperature Adjustment History:
Long press the Rotation knob >
It provides the below temperature adjustment information:
• Date and time
• Temp1 (target-actual)
• Temp2 (target-actual)
• Adjustment: done
> Adjustment > Temp adjust-Mechanical kit
> Adjustment > History-temp adjust.