EN-36
Note:
To abort Weight Adjustment, press
5. Then, the scree will display Please remove the weights.
After the weights are removed, the screen will display Adjustment in progress... >
Adjustment done. Then press tare to return to the home screen.
5.3.1.2. View Weight Adjustment History
MB62 stores the latest valid adjustment history.
To navigate to the Weight Adjustment History:
Long press the Rotation knob >
It provides the below weight adjustment information:
• Date and time
• Temperature
• Nominal weight
• Actual weight
• Difference
• Adjustment: done
5.3.2. Temperature Adjustment
Proper temperature adjustment is a critical step in ensuring accurate moisture analysis
results from your moisture analyzer. Inconsistent or inaccurate readings can occur if the
temperature is not correctly controlled during the analysis process.
The moisture analyzer adopts a two-point adjustment (100°C and 160°C). The adjustment
process takes about 30 minutes to complete.
During the adjustment, the dryer unit will heat and stabilize at the first temperature 100°C
for 15 minutes, and then at the second temperature 160°C for 15 minutes. Adjustment of
the temperature will be defined by these two points.
A Temperature Calibration Kit is required for temperature adjustment.
Note:
Mechanical Temperature Calibration Kit is available as Accessory. For details, please
refer to
Accessory (on page
5.3.2.1. Mechanical Temperature Adjustment
Prerequisites:
• Prepare a Mechanical Temperature Calibration Kit
To adjust the temperature of the moisture analyzer, processed as follows:
> Adjustment > History-Weight Adjustment.
55).