23
resonance imaging, where the intensity of EM DISTURBANCES is high.
• Use of the NE-M01 adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, NE-M01 and the other equipment should be
observed to verify that they are operating normall.
• Use of accessories, transducers and cables other than those specified or provided by the manufactu-
rer of NE-M01 could result in increased electromagnetic emissions or decreased electromagnetic im-
munity of NE-M01 and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the NE-M01, including cables
specified by the manufacturer. Otherwise, degradation of the performance of NE-M01 could result.
Note:
NE-M01 needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
The basic performance:Nebulization rate: ≥0.25 mL/min.
When the device is disturbed, the data measured may fluctuate, please measure repeatedly or in another
environment to ensure its accuracy.
Other devices may affect this device even though they meet the requirements of CISPR.
The following cable types must be used to ensure that they comply with interference radiation and immu-
nity standards:
Name
Power cord
Table 1
Guidance and Declaration - Electromagnetic Emissions
Emissions test
Radiated RF EMISSIONS CISPR 11
Radiated RF EMISSIONS CISPR 11
Harmonic distortion IEC 61000-3-2
Voltage fluctuations and flicker IEC 61000-3-3
Table 1
Guidance and Declaration - Electromagnetic Immunity
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips and
Voltage interruptions
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
IEC 60601 Test level
±8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
±2 kV for power
supply lines
± 0,5 kV, ± 1 kV, line(s) to line(s)
0 % UT; 0,5 .cycle .At0°,45°,90°,135°,180°,
225°,270°and315°. 0 % UT; 1 cycle and
70 % UT ; 25/30 cycles ;Single phase:at 0°.
0 % UT ; 250/300 cycle
30 A/m
50Hz/60Hz
ENGLISH
Cable length(m)
1,0
Compliance
Group 1
Class A
Class A
Complies
Compliance level
±8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
±2 kV for power
supply lines
± 0,5 kV, ± 1 kV, line(s) to line(s)
0 % UT; 0,5 .cycle .At0°,45°,90°,135°,180
°,225°,270°and315°. 0 % UT; 1 cycle and
70 % UT ; 25/30 cycles ;Single phase:at
0°. 0 % UT ; 250/300 cycl
30 A/m
50Hz/60Hz