Chapter 5
MAINTENANCE, TRANSPORT AND STORAGE
5.1 Clean and disinfection
Clean and disinfect the device after each use. If the device doesn't clean up, the nebulizing will be af-
fected because of drying and coagulation of medication.
1. Remove medication cup, accessory and batteries from the main unit.
2. Open the cup lid and discard residual medication.
3. Add 75% ethanol solution in medication cup, cover the cup lid, then leave for at least 10mins; it is avai-
lable to gently shake it for better disinfection.
4. Immerse the accessories to be disinfected into a container with ethanol solution, and lid the container.
Use 75% ethanol solution soaking for 10mins or longer.
5. Discard the disinfectant in medication cup, take accessories out from the disinfectant; clean the medi-
cation cup and accessories with clear-water repeatedly.
6. Fill medication cup with clear-water, assemble it to the main unit, let the device work 10mins in order
to clean the nebulizing sheet.
7. After cleaning, use new medical gauze to wipe away the water, and fully dry.
8. Use 75% medicinal alcohol to wipe the surface of main unit, then air-dry or wipe-dry with a clean, soft
cloth.
9. After all steps above, store the main unit, medication cup and accessories in a dry, clean place.
Attention:
Do not throw medication cup and accessories into boiling water for disinfection, otherwise the part may
be out of shape. Do not put them in a microwave oven for drying.
The parts disinfected with disinfectant must be fully cleaned, or the residual disinfectant may cause
symptomatic deterioration.
5.2 Medication cup replacement
The nebulizing sheet is a kind of consumable, In general, the service life of the atomizer is about six months
(20 minutes per time, three times a day).
its service life depends on the use, medication, and the degree of cleaning. If no atomizing or little ato-
mizing appears when device working, please replace medication cup in time. (If you need to purchase
medication cup, please contact the dealer.)
5.3 Transport and storage
Environment of transport and storage:
Temperature: -40°C ~ +55 ℃
Relative humidity: 5 %~96 %
Atmospheric pressure: 500 hPa ~ 1060 hPa
Requirement of transport and storage:
• No corrosion gas and well-ventilated room.
• Keep the device out of the reach of children.
• Do not store the device in places such as direct sunlight, high temperature, humid, dusty or easy to
get to water, etc.
• Avoid the device from slope, vibration or shocked.
• Transportation adopts general transportation means or follows the contract requirements. Avoid vio-
lent shock, vibration, rain and snow splash during the process of transportation.
5.4 Pollution-free disposal and recycle
The service life of product is 3 years. If the device exceeds the period of use, it must be discarded. Please
contact the manufacturer or distributor for more information.
1) The atomizer out of use can be sent back to the manufacturer or distributor for proper recycling.
2) Used parts can be returned to the manufacturer or distributor for disposal, or in accordance with rele-
vant laws and regulations.