Display Range; Limit Values; Language; Date / Time - Proceq DIGI-SCHMIDT 2000 ND Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

"0": Eliminating Individual Values
Individual values can only be deleted with this function.
• Too high or too low measured values can be eliminated
directly during the measurement process with the
PRINT key.
M1: Eliminate Highest and Lowest Value
The highest and lowest measured values are deleted
provided at least 6 measured values are present.
M3: Delete Values with Deviations of over 20 %
Values which deviate 20 % or more from the mean value
after completion of a measurement series are deleted.
The number of test impacts shown on the display is red-
uced by the number of deleted test impacts. Four beeps
are output to request replacement of the test impacts
deleted by new ones. The result of the measurement
appears on the display as soon as "n" test impacts have
been successfully performed.
Mx: Delete the Lowest and Highest Values
Only possible from EPROM version 2.0. The amount "m"
of both the lowest and highest values are deleted.
• Under "m", enter the number of measured values to be
eliminated. See also the number of test impacts "n"
set, under "Mean Value".
Example: If "m" = 3, then at least 12 should be set for "n".
© 2017 Proceq SA
4 .6

Display Range

The display range for the rebound value R on the display
can be adapted to the testing range.
• Enter R
and R
.
min
max
4 .7

Limit Values

The limit marks on the display serve to aid fast recogniti-
on of deviations from the measured values.
The range within which the rebound values R are to lie
can be selected (see Fig. 3.1 "Measuring display").
The selected limit values are not saved!
• Enter R
and R
.
min
max
4 .8

Language

• The display wording can be selected in the language
required.
In this line of the main menu "Language"
appears additionally no matter what
language is set.
4 .9

Date / Time

The inbuilt clock saves the measured values together
with the test date and time.
Correct the values set, if necessary.
Settings
11

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Digi-schmidt 2000 ld

Table des Matières