3
The lens filter indicator will blink to signal one is
needed.
To achieve accurate measurements, make sure a new,
clean lens filter is in place before each measurement.
Attach a new lens filter by pushing the thermometer
probe straight into the lens filter inside the box and
then pulling out.
Note: The Braun ThermoScan will not work unless a
lens filter is attached.
4
Select age with Age Precision
button
Press button to toggle through each age group.
NOTE: You must select an age in order to take a measurement.
The thermometer is ready to take a temperature when the display looks like the images
below.
6520
6020
Fit the probe snuggly into the ear canal, then push
5
and release the Start button
.
6
ExacTemp light will pulse while temperature measurement is in progress. The
light will remain solid for 3 seconds to indicate that a successful temperature
reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the
measurement, a long beep will sound to signal the completed measurement.
If the probe has NOT been constantly placed in a stable position in the ear
canal, a sequence of short beeps will sound, the ExacTemp light will go out
and the display will show an error message (POS = position error).
See «Errors and troubleshooting» section for more
information.
6
as shown below.
7
The confirmation beep indicates that an accurate temperature measurement has been taken.
The result is shown on the display.
0-3
3-36
36+
Months
Age range
0-3 Months
3-36 Months
36 Months -
adult
Temperature displays.
>36 Mos –
0-3 Mos
>3-36 Mos
adult
9
8
For the next measurement, press Eject button
remove and discard used lens filter, and put on a new,
clean lens filter.
NOTE: Defaults to last age setting used if you do not
change.
The Braun ThermoScan ear thermometer turns off
automatically after 60 seconds of inactivity.
The thermometer can also be turned off by
pressing the Power button
The display will briefly flash OFF and it will
go blank.
Temperature-taking hints
Always replace disposable lens filters to maintain accuracy and hygiene. The right ear
measurement may differ from the measurement taken in the left ear. Therefore, always take the
temperature in the same ear. The ear must be free from obstructions or excess earwax build-up to
take an accurate reading.
External factors may influence ear temperatures, including when an individual has:
Factor
Poor probe placement
Used lens filter
Dirty lens
0-3
3-36
36+
Months
0-3
3-36
36+
Months
Green
Yellow
Normal temperature
Elevated temperature
≥35.8 – ≤37.4 °C
(≥96.4 – ≤99.4 °F)
≥35.4 – ≤37.6 °C
>37.6 – ≤38.5 °C
(≥95.7 – ≤99.6 °F)
(>99.6 – ≤101.3 °F)
≥35.4 – ≤37.7 °C
>37.7 – ≤39.4 °C
(≥95.7 – ≤99.9 °F)
(>99.9 – ≤103.0 °F)
to
.
Affects
Red
High temperature
>37.4 °C
(>99.4 °F)
>38.5 °C
(>101.3 °F)
>39.4 °C
(>103.0 °F)