Télécharger Imprimer la page

Agilent Technologies G8703A-0 20 Notice De Mode D'emploi page 285

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 45
Pressure gauge can be connected/disconnected from the TPS-mini
during normal operation.
Figure 18
The FRG700 gauge is intended for high vacuum use, so with pressure above
NOTE
50 mTorr the signal is misinterpreted and a "false" pressure is displayed. Use a
transition gauge to understand if the reading is really under 50 mTorr or not.
Pressure Display Module
Figure 19
TPS-mini User Manual / 87-901-003-01
Gauge Connector
P R E S S U R E
x 1 0
Pressure display module
(D)
Technical Information
P1 – Vent
m B a r
T o r r
P a
285/318
17

Publicité

loading

Ce manuel est également adapté pour:

Tps-miniG8703a-021G8703a-022