Error Messages
Display Message
Cutting or grinding method not
selected
Dressing range exceeded
X-MIN sensor not activated
X-MIN sensor not deactivated
Encoder error
Cutting/grinding motor overloaded
Machine Problems
The recirculation pump is not
working correctly
Accutom-50
Instruction Manual
Explanation
 was pressed before a
START
cutting or grinding method was
selected
The parameter Y-distance is set to a
value larger than the possible
movement of the table
The inductive sensor cannot detect
the end position of the X-table
The inductive sensor for the X-posi-
tion cannot be deactivated
The Y-table does not move, or the
encoder is defective
The main motor has been exposed
to a high load for a long time
Explanation
The water level in the recirculation
tank is too low
The water outlet is clogged
The recirculation pump is
contaminated with cutting residue
Action
Select a method and press
 again
START
Adjust the Y-distance and press
START 
again
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Wait until the motor has cooled
down. "Cutting motor ready after
overload" will be shown on the
display. Adjust process parameters
to avoid repetitions
Action
Check that there is sufficient water in
the recirculation tank
Remove the cooling tubes and run
the recirculation pump.
Flush the cooling tubes with clean
water before replacing
Flush the pump with clean water by
gently pressing fresh water into the
inlet tube
57