Rotation
Rotation: Off
Rotation: Continuous
Counter-rotation
Rotation: Oscillating
Water
Starting the Cutting
22
Accutom-50
Instruction Manual
Sample rotation can be an advantage when cutting large, very hard,
coated or very long samples. On Accutom-50, three different settings
are possible.
The sample does not rotate.
The sample is rotating around its centre.
The speed can be set to three different levels, 1, 2 or 3.
The direction of rotation can be set to either counter- or co-rotation.
Counter-rotation is recommended.
Press F3 to test sample rotation with the set parameters.
Co-rotation
The sample is rocking/oscillating around its centre.
The speed can be set to three different levels, 1, 2 or 3.
The angle can be adjusted from 10° to 400°.
Press F3 to test sample rotation with the set parameters.
When the sample is moved in the x- or y- direction, with Rotation set to
Cont. or Osc. a warning message, "Rotation Mode !", will briefly flash on the
The cooling water can be set to either on or off. For all normal cutting
operations the setting should be on.
Before you start cutting make sure that the cutting chamber is clean
to ensure a free flow of cooling water. The debris collecting tray must
be free of removed material and other fragments.
Especially when using Multiple Cut, the debris collecting tray must be
cleaned before Accutom-50 is started to avoid a build-up of removed
material. This might result in a reduction of the cooling water flow, which
Position the sample correctly.
Set the correct cutting parameters.
Make sure that the correct cut-off wheel is mounted.
Close the cover of the machine.
Press START
screen as a reminder that rotation mode is selected.
IMPORTANT
might damage the sample or the cut-off wheel.
.
Note