Weekly Service
Refilling the Cooling Water
Tank
Emptying and Cleaning the Tank
Refilling the Tank
Accutom-50
Instruction Manual
Clean the specimen holder(s): movable parts, dovetail feeds and
screws. Lubricate with acid free oil.
Clean the cutting chamber and cover thoroughly. Remember to
remove all dirt particles from the grate.
Replace the cooling water in the Recirculation Cooling Unit at least
once a month.
Carefully pull out the drawer and lower the hinged front plate.
Carefully pull out the flat tank.
Remove the cover plate and empty the tank into a drain
approved for waste chemicals.
Clean the tank, the dividers and the cover with tap water. Place
the dividers correctly in the tank and replace the cover plate.
Press the cover plate into the tank and make sure it is seated
firmly.
The container and the cover plate have to be placed correctly, otherwise the
pump cannot suck up the water.
Flush the recirculation system with clean water if Accutom is not to be used
over longer periods of time. This will prevent any dried residue of cutting
material from damaging the inside of the pump.
Refill the tank with 3.88 l of water and 120 ml Struers Additive .
The water level should be 5 mm below the edge of the front hole
in the tank cover.
Push the drawer with the recirculation tank back into place.
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
IMPORTANT
IMPORTANT
IMPORTANT