Error Messages
Display Message
Searching for XY-reference position
Reference pos. not found, close
cover
Cover open!
Process stopped by open cover
Process running
Y-stop position outside cutting range
Y-stop position outside grinding
range
Y-stop position less than Y-relative
Feed less than 90% of pre-set value
Cutting stopped, feed too low
X-relative must be negative before
start
X-relative must be less than stop
pos.
Too many cuts, only xx cuts possible The no. of cuts and/or the slice
Removal too large, move X-table
56
Accutom-50
Instruction Manual
4. Trouble-Shooting
Explanation
The machine searches for the initial
reference positions
Accutom-50 was started with the
cover open and did not search for
the reference position
You are trying to start the cutting/
grinding process or a dressing while
the cover is open
You have opened the cover during a
cutting/grinding process or a
dressing process
You are trying to start two different
processes at the same time
The set stop position is outside the
maximum range in the Y-direction.
(max. 105.0 mm)
You try to start a cutting process
when the relative Y-position is larger
than or equal to the Y-stop
The force has exceeded the force
limit (LOW, MEDIUM, HIGH) and the
feed speed is automatically reduced
The feed speed has been reduced to
less than 10% due to the force
exceeding the force limit
In connection with the grinding to a
STOP POSITION the relative X-
position is positive
In connection with the grinding to a
STOP POSITION the relative X-
position is larger than the pre-set
stop position
thickness exceed the no. of possible
cuts
In connection with the grinding mode
REMOVAL, the removal is set higher
than the possible movement of the
table
Action
Wait for the search process to finish
Close the cover, wait for the
reference search to finish and press
START
again
Close the cover and press START
again
Close the cover and restart the
interrupted process
Wait for the previous process to
finish and try again
Adjust Y-stop position.
Position the specimen holder
correctly and/or change the Y-stop
position.
Decrease the feed speed or increase
the force limit next time you cut the
same material
Decrease the feed speed, increase
the force limit or change the cut-off
wheel and re-start the cutting
process
Position your sample correctly by
moving the sample holder to the right
Position your sample correctly by
moving the sample holder to the right
Re-position your sample or press
START
again to allow for
automatic adjustment of the no. of
slices
Reduce the amount of material to be
removed or clamp the sample
differently