Troubleshooting
Problem
Nothing
appears on
the display,
even when
the power is
turned on.
The cuff does
not inflate.
The device
does not
measure.
Readings are
too high or too
low.
Other
Note: If the actions described above do not solve the problem, contact the dealer. Do not attempt to
open or repair this product, as any attempt to do so will make your warranty invalid.
E-26
Probable Cause
Battery is drained.
Battery terminals
are not in the correct
position.
Battery voltage is
too low.
(LOW
BATTERY mark) blinks.
If the batteries are
drained completely, the
mark does not appear.
The cuff is not applied
properly.
Readings are too high
or too low.
The cuff position is not
correct.
______________
The value is different
from that measured
at a clinic or doctor's
office.
______________
Corrective Action
Replace all batteries with new ones.
Reinstall the batteries with negative
and positive terminals matching those
indicated on the battery compartment.
Replace all batteries with new ones.
Apply the cuff correctly.
Make sure you remain very still and
quiet during measurement.
Sit comfortably and still. Place your arm
on a table with your palm facing upward
and the cuff at the same level as your
heart.
If you have a very weak or irregular heat
beat, the device may have difficulty in
determining your blood pressure.
Refer to "Why Measure Blood Pressure
at Home?".
Remove the batteries. Place them
back properly and take another
measurement.