Measurements
Measurement with Set Pressure
The UA-1030TCN is designed to detect the pulse and to inflate the cuff to
a systolic pressure level automatically. If re-inflation occurs repeatedly, use
the following methods. During the blood pressure measurement, re-inflation
may occur. A fixed pressure value can be set to avoid re-inflation.
1. At step 5 of the clock adjustment
procedure (see page E-12), press the
set button to go to the pressure setting
mode. The current setting blinks.
2. Press the
pressure value about 30mmHg or more
above your expected systolic pressure
from the following.
AUTO
: Automatic pressurization (default value)
180
: Pressure value of 180mmHg (fixed)
210
: Pressure value of 210mmHg (fixed)
240
: Pressure value of 240mmHg (fixed)
3. Press the set button to go to the TriCheck setting mode.
Press the START button to turn the device off. After three minutes of
non-operation, the device will turn off automatically. The next
measurement will be performed with the new pressure value.
or
button to select a
or
TriCheck
Turn off
setting mode
At heart level
Press and hold the button
Release the button
at the desired
systolic pressure
Refer to E-17
for measurement
START
E-19