Using the Monitor
Selecting the Correct Cuff Size
Using the correct cuff size is important for an accurate reading. If the
cuff is not the proper size, the reading may yield an incorrect blood
pressure value.
• The arm size is printed on each cuff.
• The index
and proper fit range, on the cuff, tell you if you are applying the
correct cuff. (Note: the SmoothFit Cuff, UA-392A, does not have the index
symbol)
• If the index
purchase a replacement cuff. The arm cuff is a consumable. If it becomes
worn, purchase a new one.
Catalog Number
UA-289A
UA-290A
UA-291A
UA-392A
UA-420A
Applying the arm cuff
1. Wrap the cuff around the upper arm,
about 1-2 cm above the inside of the
elbow, as shown in the figure to the right.
2. Place the cuff directly against the skin.
Clothing may cause a faint pulse,
and result in a measurement error.
Constriction of the upper arm, caused
by a tightly rolled up shirtsleeve, may
prevent accurate readings.
points outside of the range, contact your local dealer to
Recommended Cuff Size
Small Cuff
Medium Cuff
Large Cuff
SmoothFit
Wide Range Cuf
Cuff
TM
Arm Size
16-24cm
23-37cm
31-45cm
23-37cm
22-42 cm
Do not roll up shirtsleeve tightly
Cuff
1-2 cm
Artery position mark
Secure the fabric
fastener firmly
Index
and proper fit range
Air Hose
E-15