Using the Programming Menus
•
Press enter to confirm and the screen displays the
concentration during calibration.
A success or failure message displays and the screen returns to the
main menu option.
TestMode Menu
The TestMode menu performs a test of all functions on the unit. All
outputs are activated according to their failsafe mode, relay setting, and
buzzer setting. All functions are activated simultaneously for a maximum
of 3 minutes (or until the user cancels the test) and the 4-20mA output
is saturated at 20 mA to ensure working order. The unit does not
detect gas during this TestMode procedure. Press enter to access the
TestMode.
•
Use the arrow keys to scroll to Yes (or No) and press any key to
begin the test sequence.
•
To cancel the test, press enter at anytime when Stop displays.
24
Set Span
198 ppm
* Menu *
TestMode
TestMode
Yes
TestMode
Stop
E
Point Standalone Gas Monitor User Manual
3