Using the Programming Menus
Press enter to confirm and the screen displays the
concentration during calibration.
A success or failure message displays and the screen returns to the
main menu option.

TestMode Menu

The TestMode menu performs a test of all functions on the unit. All
outputs are activated according to their failsafe mode, relay setting, and
buzzer setting. All functions are activated simultaneously for a maximum
of 3 minutes (or until the user cancels the test) and the 4-20mA output
is saturated at 20 mA to ensure working order. The unit does not
detect gas during this TestMode procedure. Press enter to access the
TestMode.
Use the arrow keys to scroll to Yes (or No) and press any key to
begin the test sequence.
To cancel the test, press enter at anytime when Stop displays.
24
Set Span
198 ppm
* Menu *
TestMode
TestMode
Yes
TestMode
Stop
E
Point Standalone Gas Monitor User Manual