Using the Programming Menus
•
The Hyst(eresis) screen displays. This value, added to the alarm
level, tells the unit at which level to deactivate the alarm
(e.g. Alarm level A is 25 ppm, minus a hysteresis level of -05,
gives a deactivation level of 20 ppm). Use the arrow keys to
increase or decrease the value and press enter. A confirmation
screen displays. Press enter to confirm.
The screen briefly displays a success message and returns to the main
menu option.
Restore Menu
CAUTION:
!
Restore function of this unit. Factory calibration settings conform to specific
standards. Any changes made may affect manufacturer's stated standards
compliance.
The Restore menu allows users to restore all calibration information to
factory configured settings. Press enter to access the Restore menu.
•
Use the arrow keys to scroll to Yes (or No) and press enter to
select.
The screen briefly displays a success message and returns to the main
menu option.
20
SetAlarm
Only qualified, knowledgeable personnel should use the
* Menu *
Restore
Restore
Restore
Success
Yes
Yes
E
Point Standalone Gas Monitor User Manual
3