specified, may affect the long term reliable operation of the product.
DANGER
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned on but not in use, the oxygen will make the materials flammable. Turn the oxygen concentrator OFF when not in use to prevent oxygen enrichment.
NOTE– If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
NOTE– If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Troubleshooting chart on page 10, and contact your
DeVilbiss provider if necessary.
2.
Check the flow meter to make sure that the flow meter ball is centered on the line next to the prescribed number of your flow rate.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
NOTE– Your Drive DeVilbiss provider may have preset the flow meter so that it cannot be adjusted.
NOTE– If the flow meter knob is turned clockwise, the flow decreases (and eventually will shut off the oxygen flow). If the knob is turned counterclockwise, the flow
increases.
NOTE– For prescriptions of 10 LPM, be sure the ball is centered on the 10 liter line. The ball should not touch the red line. Setting the flow higher than 10 may cause
the oxygen purity level to drop.
NOTE– The low-flow alarm may activate if the flow meter ball is set below 2 lpm. The unit will continue to run; however, the Service Required light will come on
accompanied by an audible alarm. Adjust the flow meter to your prescribed flow.
IMPORTANT
Do not use a low-output flow meter with this concentrator.
3 .
Your Drive DeVilbiss concentrator is now ready for use, properly position the cannula with the nasal prongs facing upward, insert the
prongs into nose. Wrap the cannula tubing over the ears and position in front of body (Figure 4). Allow 15 minutes for the oxygen
concentrator to reach stated performance .
®
DeVilbiss OSD
Operation
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit.
The OSD lights on the top panel are defined as follows:
Green Normal Oxygen light–acceptable oxygen level .
Yellow Low Oxygen light– below an acceptable oxygen level .
If the oxygen purity falls below the acceptable level: The green Normal Oxygen light will shut off, the yellow Low Oxygen light will illuminate, and an intermittent audible signal
will sound .
Refer to the Troubleshooting section in this guide on page 10, and switch to your reserve oxygen system . Do not attempt any other maintenance . Contact your Drive DeVilbiss
provider immediately .
RESERVE OXYGEN SYSTEM
As a precaution, your Drive DeVilbiss provider may supply you with a reserve oxygen system . If your unit loses electrical power or fails to operate correctly, the Patient Alert
System will sound to signal you to switch to your reserve oxygen system (if provided) and contact your Drive DeVilbiss provider. Please contact your oxygen provider if you
have questions regarding a reserve oxygen system .
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR
Drive DeVilbiss recommends using only original DeVilbiss parts and filters in order to guarantee reliable operation of the product.
WARNING
Do NOT use lubricants, oils or grease .
Before attempting any cleaning procedures, turn the unit "
Cannula/Mask, Tubing and Humidifier Bottle
Clean and replace the cannula/mask, tubing and humidifier bottle according to the manufacturer's instructions.
Filter Door with Vents
Inspect the vents periodically, and wipe with a dry cloth as needed to remove dust .
Exterior Cabinet
Clean the concentrator exterior cabinet weekly by using a damp cloth and wiping dry; the vents can also be wiped with a damp cloth .
Cleaning
Outer Cabinet
Filter Door Vents
Oxygen Outlet Connector
* number of cleaning cycles determined by recommended cleaning interval and expected service life
A-1025DS
OFF
."
Recommended cleaning
interval
7 days
7 days
7 days
Number of cleaning
cycles *
260
260
104
4
Compatible cleaning method
Water, use only a damp cloth
Wipe with dry cloth, or a cloth dampened with
water to remove dust .
Mild dish soap (2 tbsp) and warm water (2 cups)
EN - 9