OVERVIEW OF ALARMS AND SERVICE INDICATORS
This device contains an alarm system which monitors the state of the device and alerts of abnormal operation, loss of essential performance or failures . Alarm conditions are
shown on the LED display. The alarm system functions are tested at power up by lighting all visual alarm indicators and sounding the audible alarm (beep).
All alarms are Low Priority Technical Alarms .
Alert or Alarm
LED Icon
Condition
Start-up Period
Low Oxygen Output
Concentration
Device Malfunction
SPECIFICATIONS
Catalog Number
Delivery Rate
Maximum Recommended Flow (@
nominal outlet pressures of zero and
7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Atmospheric Pressure
1010 hPa to 840 hPa
0-1500 M (0-4921 ft)
Operating Environment Range
Operating Relative Humidity Range
Power Consumption
Weight
Safe Working Load
Sound Pressure Level at 3 and
10LPM
Sound Power Level at 3 and 10 LPM
Dimensions
Maximum Limited Pressure
Operating System
Low Oxygen Indicator
Storage Conditions
Equipment Class and Type
A-1025DS
Details of Alert or Alarm
Condition
The unit has recently been
started and is in start-up
period, the output flow of the
oxygen is temporarily < 82%
The output flow of oxygen is
≤ 82%, which indicates the unit
may need routine servicing
indicating a Low O2 condition
The device is experiencing a
malfunction that requires
servicing to correct
DEVILBISS 10-LITER SERIES
1025DS
2 to 10 LPM
10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
120 V, 60 Hz, 6 .1 Amp
102-132 V~, 60 Hz
2-10 LPM=87%-96%
Across the voltage range; tested at 70˚F
(21˚C) only; No degradation of
performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
120 vac, 60HZ: 639 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<59 dBA
<69 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IP21
Visual Alert or Alarm
The YELLOW LED light on
No audible alarm during
the panel is illuminated
indicating low O2 condition
The YELLOW LED Light on
The audible alarm is beeping
the panel is illuminated,
The RED Service Required
The audible alarm is beeping
LED light is illuminated
1025KS
2 to 10 LPM
10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
230 V~, 50 Hz, 3 .2 Amp
195-253 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range; tested at 70˚F
(21˚C) only; No degradation of
performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
230 vac, 50Hz: 664 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<57 dBA
<67 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
IP21
Audible Alarm
Wait for unit to finish start-up
start-up period
period, up to 15 minutes
Contact your Oxygen Equipment
intermittingly
Provider for assistance and to
arrange for servicing of the unit
Contact your Oxygen Equipment
intermittingly
Provider for assistance and to
arrange for servicing of the unit
1025UK
2 to 10 LPM
20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
240 V~, 50 Hz, 3 .2 Amp
204-264 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range; tested at 70˚F
(21˚C) only; No degradation of
performance
41˚F (5˚C) to 95˚F (35˚C)
15% to 93%, non-condensing
240 vac, 50Hz: 670 watts average
42 lbs. (19 Kilograms)
53 lbs. (24 Kilograms)
<57 dBA
<67 dBA
24 .5"H x 13 .5"W x 12"D
(62.2 x 34.2 x 30.4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
<82% low oxygen
<60% very low oxygen
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Class II Equipment Double Insulated;
Type B Applied Part
Action
10 LPM
IP21
EN - 11