Next screw on the Filling Chamber Cap
again.
Screw on Filling Chamber Cap
Dronabinol (THC) is available as an al-
coholic solution.
The advantage of this is that the liquid
is distributed more equally around the
stainless steel wire of the Filling Pad,
yielding a larger surface area that can
be vaporized.
However, the alcohol must first be
separated from the dronabinol, as
alcohol should not be inhaled.
To do that, place the Filling Chamber
with cap on the Hot Air Generator and
screw it on clockwise as far as it will
go. Switch on the heater (HEAT), adjust
the set temperature to 100°C (212°F)
and let it heat up until the actual tem-
perature reaches 100°C (212°F). Then
press the touch button labeled ‚AIR' to
pump air through the Filling Pad. The
dronabinol does not vaporize at this
temperature, but the alcohol evapo-
rates quickly (within 30 sec.).
6. U S A G E O F D R O N A B I N O L
Since alcohol has a characteristic
odor, you can easily determine by
smelling whether the Filling Chamber
is alcohol free. Dronabinol has neither
odor nor taste.
Separate alcohol
When the Filling Chamber is free of al-
cohol, press the touch button labeled
‚AIR' again to switch off the airflow.
The dronabinol, which evaporates only
at higher temperatures - over 157°C
(315°F), remains in the Filling Pad.
Then adjust the set temperature to
210°C (410°F) and continue heating.