5. T H E V O L C A N O M E D I C 2 H O T A I R G E N E R AT O R
5.6. Temperature Setting
Actual Value Display
(orange)
Mains Operation
Display
Touch Button
Temperature Minus
Touch Button
Heater (HEAT)
When you briefly tap on the touch but-
ton Temperature Plus or Minus, the set
values are increased or decreased in
the respective direction in 1-degree in-
crements. If you keep the touch button
Temperature Plus or Minus pressed,
the set values start running conti-
nuously in the corresponding direc-
tion. If the touch button Temperature
Plus and Minus is activated simulta-
neously, the display changes from de-
gree Celsius to degree Fahrenheit and
vice versa.
If the actual value is higher
than the set value, you can
speed up the cooling down
process by switching on the air
pump.
Pump Display
30 minutes after the last time a
button is pressed, the device
automatically switches off.
As long as the heater has a substantial
amount of residual warmth, the tempe-
rature is displayed until it drops below
+40°C (+104°F).
Then the display turns off, and the on
the mains operation display indicates
that the Hot Air Generator is connected
to the mains supply.
The device is safely discon-
nected from the mains only
when the plug is removed
from the socket.
92
Set Temperature
Display (white)
Touch Button
Temperature Plus
Touch Button
Pump (AIR)