9.1. General
If the VOLCANO MEDIC 2 Vaporization
System is to be used by other people
(patients), the Filling Chamber must be
cleaned and a new Valve Balloon with
Mouthpiece or a new Tube Kit with
Mouthpiece must be used. Further
information on the Filling Chamber,
Valve Balloon and Tube can be found
in Chapter 9.3. „Filling Chamber, Tube
Flexer and Mouthpiece" are described
from page 116 onwards.
Valve Balloon with Mouthpiece and
Tube Kit with Mouthpiece may only be
used by one person.
Caution!
To avoid a health risk, e.g. an
infection from a contamina-
ted Filling Chamber, Valve Balloon or
Tube Kit, the following hygiene instruc-
tions must be followed.
The Valve Balloon or Tube Kit may only
be used by one person.
Make sure that the equipment is al-
lowed to dry thoroughly after every
cleaning. Condensation or residual
moisture may encourage bacterial
growth.
Please note Chapter 9.2. „Material Re-
sistance of the VOLCANO MEDIC 2 Va-
porization System", page 116.
9. C L E A N I N G / H Y G I E N E
113
Inspect the component parts of the
Filling Chamber regularly and replace
defective (broken, misshapen, discolo-
red) parts. Replace the Valve Balloon
or Tube Kit after no more than 70 ap-
plications or at least 14 days after the
first use.