Push the lower Screen upwards out of
the Filling Chamber Housing.
Push the lower Screen out
9.4.4. Drying
All component parts of the Filling
Chamber must be completely dry be-
fore reassembly.
Put all component parts of the Filling
Chamber on a clean, dry, and absor-
bent pad and allow them to dry tho-
roughly. A hair dryer may be used for
faster drying. Do noy dry in wet rooms
(e.g. not in the bathroom).
9. C L E A N I N G / H Y G I E N E
Lower Screen
Filling Chamber
Cylinder
Filling Chamber
Housing
121
9.4.5. Assembly of the Filling
Chamber Cap
Assemble the components in the re-
verse order: Slide the cap cylinder into
the cap housing until it clicks, then pull
the cap ring from above over the cap
cylinder to the cap housing.
When using hemp flowers, push the
upper Screen from below into the cap
cylinder until it clicks into the designa-
ted notch.
When using dronabinol solution, the
two Screens are not needed. Only the
Filling Pad is inserted instead of the
lower Screen.
Cap Ring
Cap Housing
Touch Guard
Cap Cylinder
Upper Screen