Immunity test
Conducted RF
IEC 61000-4-6
Amplitude
modulated
Radio-frequen-
cy electroma-
gnetic field
Amplitude
modulated
IEC 61000-4-3
10. H O T A I R G E N E R AT O R
IEC 60601 test
level
3 V at 150 kHz-
80 MHz
6 V in ISM and
amateur radio
bands
80 % AM at 1
kHz
10 V/m at
80 MHz -
6000 MHz
Home
Healthcare
Prof. Healtcare
80 % AM at 1
kHz
Compliance
level
3 V at 150 kHz-
80 MHz
6 V in ISM and
amateur radio
bands
80 % AM at 1
kHz
10 V/m at
80 MHz -
2.7 GHz
Home
Healthcare
Prof. Healtcare
80 % AM at 1
kHz
132
Electromagnetic
environment – guidance
WARNING: Portable and
mobile RF communica-
tions devices (including
their accessories such
as external antennas
and antenna cables)
should be used no
closer than 30 cm (12 in-
ches) to any part of the
VOLCANO MEDIC 2
Vaporization System, in-
cluding cables, specified
by the manufacturer.
Otherwise, degradation
of the performance of
this equipment could
result.
WARNING: Portable and
mobile RF communica-
tions devices (including
their accessories, such
as external antennas
and antenna cables)
should be used no
closer than 30 cm (12 in-
ches) to any part of the
VOLCANO MEDIC 2
Vaporization System, in-
cluding cables, specified
by the manufacturer.
Otherwise, degradation
of the performance of
this equipment could
result.