Télécharger Imprimer la page

I-Tech MNPG166 Mode D'emploi page 44

Publicité

Tableaux EMC
EMC Tabels
Emission test
RF emissions
Cispr 11
RF emissions
Cispr 11
The device is inteded for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that is used in suche environment
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Mains
power
electromagnetic field
EN 61000-4-8
IACER S.r.l.
Electromagnetic emission
Compliance
Group 1
Class B
Electromagnetic immunity
Test level EN
Compliance
60601-1-2
 6kV contact
 6kV contact
 8kV air
 8kV air
3 A/m
3 A/m
44 di 48
Electromagnetic environment –
guidance
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Electromagnetic
environment – guidance
level
Floors sholud be wood,
concrete or ceramic tile. If
floor are covered with
syntethic material, the
relative humidity should be
at least 30%
Mains power quality should
be at that of a typical
commercial or hospital
environment
MNPG166 Rev. 05 27/06/16