5. Cleaning and maintenance

The Elexo XXL must be cleaned after every use, with water and regular household
cleaning agents. PUR integrated foam components with IMC coating should be rubbed
clean and disinfected. Do not use any abrasive or other aggressive cleaning products.
If necessary, the PUR components can be disinfected with a solution of 70% ethanol or
propanol.
The time the products must be left on the components has a significant impact upon the
lifespan of the PUR components. As a result of the material properties of PUR, contact
with sharp objects must be avoided. This could lead to damage or tears in the material.
Finally, we would like to point out that non-critical parts of the PUR components must
also be cleaned and disinfected.
Check the wheels regularly for hairs, dust and soap residues and remove when necessary.
Maintenance and repairs to the Elexo XXL may only be carried out by authorised personnel.