Check whether the danger has passed or the problem has been resolved before
resetting the emergency stop.
The orange LED light on the manual control will flash when the emergency
stop has been activated.
4.2. Sleep setting
After around an hour, the chair will fall into a 'sleep' setting and the battery indication LEDs
on the manual control will go out. In order to use the chair again, it must be activated
(See paragraph 2.2. Activating).
4.3. Emergency lowering
If the controls stop working and the chair is in a raised position,
it can be lowered using the emergency lowering button.
Check first to make sure the chair is not in sleep setting!
• Remove the plug from the manual control (1) (See figure 18).
• Press the emergency lowering button (2) (See figure 18).
The chair will now return to the lowest position (make sure
no persons or objects are under the chair). Take the client from the chair and make
sure the device is not used until it has been repaired. Place the chair in a safe place and
indicate that it is out of order.
Contact the technical service or Lopital's service department. E-mail: service@lopital.nl
4.4. On / off switch
On the right hand side of the cover cap, there is the on / off
switch (See figure 19). If the chair is out of use for an extended
period, you are advised to turn it off. This will prevent any
unauthorised use of the chair.
Elexo XXL Shower-toilet chair | Instructions for use | 37
2
1
Figure 18 Emergency lowering
button.
Figure 19 On / off switch.