ASSEMBLY INSTRUCTIONS (CONTINUED)
Assembling the Patient Lifter
1. Place the Stretcher Storage Tray on the Lifter Base, as shown below,
before attaching the Lifter Mast.
2.
Remove black plastic protective caps from Lifter Mast. Insert the bottom of the Lifter Mast
and Hydraulic Pump assembly into the mast sleeve (through the Stretcher Tray Holder) on
the Lifter Base. The notch on the end of the mast fits over the round tube at the bottom of
the Mast Sleeve. The mast must lock into position in the mast sleeve, making it impossible
to rotate. Check that the mast is locked into position and that the hole in the mast is
aligned with the threaded hole in the mast sleeve.
3.
Insert the threaded locking device (the plastic knob and stud chained to the base) into the
threaded hole in the mast sleeve. Tighten to lock the mast to the base. Keep mast and
base locked at all times except when removing mast from base for storage.
Figure 2
- 7 -