2 Installing the Instrument
4. Attach the safety cover to the rear panel. Make sure that you do not pinch the sense wires when
attaching the cover. Note that heavy wiring cables must have some form of strain relief to prevent
bending the safety cover or bus bars.
Tightening torque for the 8.8 mm opening cannot exceed 10.8 Nm (8 lb-ft).
Tightening torque for the 5.5 mm opening cannot exceed 4.8 Nm (3.5 lb-ft).
Ferrite Core Installation - for PV8922A, PV8931A, PV8932A models only
For the applicable units, install the supplied ferrite core to
reduce EMI (electromagnetic interference).
Run each output lead through the core three times. If using
remote sensing (not shown), also run each sense lead
through the core three times.
Place the core as close as possible to the output cover as
shown.