DISPlay Subsystem
Display commands control the front panel display.
DISPlay[:WINDow][:STATe] 0|OFF|1|ON
DISPlay[:WINDow][:STATe]?
Turns the front panel display on or off.
Parameter
0|OFF|1|ON, *RST ON
Turns the front panel display off: DISP OFF
DISPlay[:WINDow]:VIEW METER_VI|METER_VP|METER_VIP
DISPlay[:WINDow]:VIEW?
Selects the parameters to display on the front panel. 
METER_VI displays output voltage and current.
METER_VP displays output voltage and power.
METER_VIP displays output voltage, current, and power.
Parameter
METER_VI|METER_VP|METER_VIP,
*RST METER_VI
To display voltage and power: DISP:VIEW METER_VP
DISPlay:SAVer[:STATe] 0|OFF|1|ON
DISPlay:SAVer[:STATe]?
Turns the front panel screen saver on or off.
Parameter
0|OFF|1|ON, *RST OFF
Turns the front panel screen saver on: DISP:SAV ON
Keysight PV8900 Series Operating and Service Guide
Typical Return
0 or 1
Typical Return
METER_VI, METER_VP,
or METER_VIP