Télécharger Imprimer la page

Keysight PV8900 Serie Guide D'utilisation Et De Maintenance page 246

Simulateurs de groupe photovoltaïque

Publicité

Les langues disponibles

Les langues disponibles

5 SCPI Programming Reference
SCPI Error Messages
The Keysight instrument returns error messages in accord with the SCPI standard.
Up to 20 errors can be stored in each interface-specific error queue (one each for GPIB, USB, VXI-
l
11, and Telnet/Sockets.) Errors appear in the error queue of the I/O session that caused the error.
The front-panel ERR annunciator turns on when there are one or more errors in the error queue.
l
A special global error queue holds all power-on and hardware-related errors (i.e. power-fail).
l
Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. Once you have
l
read all interface-specific errors, the errors in the global error queue are retrieved. When you have
read all errors from the error queue, the ERR annunciator turns off.
If more than 30 errors have occurred, the last error stored in the queue (the most recent error) is
l
replaced with -350,"Error queue overflow". No additional errors are stored until you remove errors
from the queue. If no errors have occurred when you read the error queue, the instrument
responds with +0,"No error".
The front panel reports errors from all I/O sessions and the global error queue. To display the error
l
queue on the front panel, press the Error key. Displaying the errors clears the error queue.
Error conditions are also summarized in the Status Byte Register. See
l
duction
for details
The interface-specific error queues are cleared by power cycles and *CLS. The error queue is not
l
cleared by *RST.
SCPI: SYSTem:ERRor?
l
Errors have the following format (the error string may contain up to 255 characters):
-113 "Undefined header".
Error Device-dependent Errors (these errors set Standard Event Status register bit #3)
0 No error
This is the response to the ERR? query when there are no errors.
101 Calibration state is off
Calibration is not enabled. The instrument will not accept calibration commands.
102 Calibration password is incorrect
The calibration password is incorrect.
103 Calibration is inhibited by switch setting
Calibration mode is locked out by the calibration switch.
104 Bad sequence of calibration commands
Calibration commands have not been entered in the proper sequence.
246
Read and clear one error from the queue
Keysight PV8900 Series Operating and Service Guide
Status Subsystem Intro-