6 Verification and Calibration
Verification Procedure
Voltage Programming and Readback Accuracy
This test verifies that the voltage programming and measurement functions are within specifications.
Step 1. Turn off the unit under test. Only connect the DMM to the sense screws (see
Test Setup
A). Do
NOT connect the RP79xxA load unit.
Step 2. Turn on the unit under test and program the instrument settings as described in the test record
form under "Voltage Programming & Readback, Min Voltage". Turn the output on. The output status
should be "CV", with the output current close to zero.
Step 3. Record the output voltage reading from the DMM (Vout) and the voltage measured by the
instrument over the interface. The readings should be within the limits specified in the test record form
for the appropriate model under "Voltage Programming & Readback, Minimum Voltage".
The next step applies the full output voltage of the unit under test to the output.
Step 4. Program the instrument settings as described in the test record form under "Voltage
Programming & Readback, High Voltage".
Step 5. Record the output voltage reading from the DMM (Vout) and the voltage measured by the
instrument over the interface. The readings should be within the limits specified in the test record form
for the appropriate model under "Voltage Programming & Readback, High Voltage".
Constant Voltage Load Effect
This test measures the change in voltage resulting from a change in current from full load to no load.
Step 1. Turn off the unit under test. Connect the DMM to the sense screws and connect the RP79xxA
load unit (see
Test Setup
A).
Step 2. Turn on the unit under test and program the instrument settings as described in the test record
form under "CV Load Effect". Turn the output on.
Step 3. Set the RP79xxA load for the rated output current of the unit under test as described in the test
record form under "CV Load Effect". Turn the output on. The output status of the unit under test should
be "CV". If it isn't, adjust the load so that the output current drops slightly.
Step 4. Record the output voltage reading from the DMM.
Step 5. Turn off the output of the RP79xxA load. Record the voltage reading from the DMM again. The
difference between the DMM readings in steps 4 and 5 is the load effect, which should not exceed the
value listed in the test record form for the appropriate model under "CV Load Effect".
Constant Voltage Ripple and Noise
Periodic and random deviations in the output combine to produce a residual ac voltage superimposed
on the dc output voltage. This residual voltage is specified as the rms or peak-to-peak noise in the
indicated frequency range (refer to the
Specifications
section).