• The settings of other models or brands of oxygen
therapy equipment may not correspond with the
settings of this device.
• The settings of this device may not correspond
with the settings for devices that provide a
continuous flow oxygen.
• Use of this device at an altitude above 3,048 m
(10,000 ft) or outside the temperature range of
5 – 40°C (41 – 104° F) or a relative humidity above
95% is expected to adversely affect the flowrate
and the percentage of oxygen and consequently
the quality of the oxygen therapy. Use of
this device immediately after storage in
temperatures beyond the allowable operating
range may adversely affect operation of the
device until the temperature returns to the
allowable operating range. Wind or strong drafts
can adversely affect the accurate delivery of
oxygen therapy.
• If the device fails, it will cause a return to your
previous condition prior to starting oxygen
therapy. This state will be different for each
individual patient.
• If you are unable to communicate discomfort,
you may require additional monitoring and or a
distributed alarm system to convey the
information about the discomfort and or the
medical urgency to your responsible
caregiver to avoid harm.
4.2 CAUTION
Risk of minor injury or discomfort
• Use of this device has not been studied in
pediatric populations. Consult your physician
before using the product for pediatric patients.
• Incompatible parts and accessories can result in
degraded performance or damage and may void
your warranty.
• The device is designed to provide a flow of high
purity oxygen. An advisory alert, "Oxygen Low",
will inform you if oxygen concentration drops.
If alarm persists, contact your equipment
provider.
• The oxygen flow setting must be determined and
recorded for each patient individually by the
prescriber, including the configuration of the
device, its parts and the accessories. It is
the patient responsibility to check with
provider to reassess settings of the therapy for
effectiveness.
©2024 Inogen, Inc. All rights reserved.
• It is the responsibility of the patient to plan for
a back-up oxygen supply when traveling;
Inogen assumes no liability for any
disruptions in oxygen supply if a backup
source is not secured.
• It is the responsibility of the patient to use only
parts and accessories mentioned in these
instructions for use. Parts and accessories
used by the patient not recommended in these
instructions for use are at the sole responsibility
for the patient. Inogen assumes no liability for
use of parts and accessories not mentioned in
these instructions for use.
• It is the responsibility of the patient to
periodically check the battery and replace as
necessary per these instructions for use. Inogen
assumes no liability for persons choosing not to
adhere to manufacturer recommendations.
• Do not modify the device. Incompatible parts
and accessories as a result of modifications
can degrade performance or cause damage
and may void your warranty unless indicated
or instructed to do so.
• Do not use this product in any way other than
described in the specifications and intended use
sections of this manual as it may lead to product
damage, loss of product function, or personal
injury.
• Do not obstruct air intake or exhaust when
operating the device. Blocking air circulation
or placing close to a heat source may lead to
internal heat buildup and shutdown or damage to
the concentrator. In the event of changes to the
performance of the device, please refer to the
troubleshooting section of this document.
• Do not operate the device without the particle
filter in place. Particles drawn into the system
may damage the equipment.
• Do not wrap cords around power supply for
storage. Do not drive, drag or place objects over
cord. Doing so may lead to damaged cords and a
failure to provide power to the concentrator.
• Do not use the DC power cord with a plug
splitter. This may cause overheating of the DC
power cord.
• Do not disassemble the power supply. This may
lead to component failure and/or safety risk.
• Do not place anything in the device's power
port other than the supplied power supply. If an
extension cord is used, use an extension cord
Page 7 of 120
96-13115-50-01 B