Display Icon
©2024 Inogen, Inc. All rights reserved.
Description
Oxygen delivery error: A breath
has been recognized, but proper
oxygen delivery has not been
detected.
Battery empty, attach plug:
The concentrator has insufficient
battery power. The concentrator
will shut down and stop
producing oxygen.
Battery HOT : The battery has
exceeded temperature limit
while concentrator is running on
battery power. The concentrator
will shut down and stop
producing oxygen.
System HOT: Concentrator
temperature is too high. The
concentrator will shut down and
stop producing oxygen.
Sensor fail: The
concentrator's oxygen
sensor has malfunctioned.
System COLD : The system is
cold (<2˚C). The concentrator will
shut down and stop producing
oxygen.
System Error: The concentrator
will shut down and stop
producing oxygen.
Page 24 of 120
What To Do
If condition persists, switch to
backup oxygen source and
contact your equipment provider
to arrange for service.
Attach an external power supply
or replace with a fully charged
battery. If the device has turned off,
press and hold the power button to
turn back on.
If possible, move concentrator to
a cooler location, then turn power
off and back on. Ensure air intake
and outlet vents have clear access
and particle filters are clean. If
condition persists, switch to
external power or a backup source
of oxygen and contact your
equipment provider.
Ensure air intake and outlet vents
have clear access and particle
filters are clean. If condition
persists, switch to a backup source
of oxygen and contact your
equipment provider.
You may continue to use the
concentrator. If the condition
persists, contact your equipment
provider.
Move to a warmer environment to
allow the unit to warm up before
starting it. If condition persists,
switch to a backup source of
oxygen and contact your
equipment provider.
Switch to backup oxygen source
and contact your equipment
provider.
96-13115-50-01 B