10.6 BATTERY CARE AND MAINTENANCE
Lithium-ion batteries require special care to ensure proper performance and long life. Use only compatible
batteries with your device.
• Keep Dry: Always keep liquids away from batteries. If batteries become wet, discontinue use
immediately and dispose of battery properly.
• Effect of temperature on battery performance: The battery powers the device under most
environmental conditions. To extend the run-time of your battery, avoid running in temperatures
less than 41˚F (5˚C) or higher than 95˚F (35˚C) for extended periods of time.
• Battery Storage: Remove your battery from the device when it is not in use to avoid inadvertent
discharge. Store battery in a cool, dry place. Store with a charge of at least 40-50%. Batteries
should be charged up to full charge and discharged down to 0% at least once every 90 days to
maintain maximum lifetime. Avoid storing your device Battery in extreme temperatures, below -4˚F
(-20˚C) or above 140˚F (60˚C), for any amount of time.
• Battery Disposal: Batteries must only be placed in the collection containers for waste portable
batteries when they are discharged, or when precautions against short circuits have been taken in
the case of batteries that are not completely discharged (e. g. by isolating the poles with adhesive tape).
Lithium-ion batteries, like all rechargeable batteries, are recyclable and should never be incinerated.
10.7 DC POWER CABLE FUSE REPLACEMENT (RP-125)
The DC power cable contains a fuse. If the DC power cable is being used with a known good power source
and the device is not receiving power, the fuse may need to be replaced.
To replace the fuse:
1. Remove the tip by unscrewing the retainer. Use a tool if
necessary.
2. Remove the retainer, tip, and fuse.
3. The spring should remain inside the adapter housing.
4. If the spring is removed, replace the spring first before
inserting the replacement fuse.
5. Install a replacement fuse
6. Reassemble the tip.
7. Ensure the retainer ring is properly seated and tightened.
WARNING
• CHOKING HAZARD: small parts exposed when changing the fuse, keep away from small children
and pets.
• CRITICAL FUSE SIZING: incorrect fuse replacement size may result in fire or inadequate equipment
protection. Replace only with same type and rating of fuse.
• ELECTRICAL SHOCK: completely disconnect the cable before attempting to change the fuse.
• Do not hang any type of accessory or accessory bracket from plug or cable.
11. DEVICE REPAIR & DISPOSAL
11.1 REPAIR
Do not attempt to repair the device unless otherwise specified in these instructions for use. Contact your
equipment provider or Inogen for assistance.