Télécharger Imprimer la page

Inogen Rove 6 Manuel De L'utilisateur page 27

Masquer les pouces Voir aussi pour Rove 6:

Publicité

Les langues disponibles

Les langues disponibles

Periodic visual inspection of the device is required to ensure no damage to the exposed components is
apparent. A typical visual inspection includes:
• Battery connectors - these should not be bent or deformed.
• Cannula barb - this should be straight and fully seated against the housing.
• Housing - the housing should be fully seated and secure with no cracking or other visible damage.
• Particle filters - these should be in place and clear of debris, dust or other obstructions.
Replacement parts can be purchased from your equipment provider or the manufacturer Inogen,
at Inogen.com or by calling 1-877-466-4364.
10.1 CANNULA REPLACEMENT
Your nasal cannula should be replaced on a regular basis per the manufacturer's instructions for use.
Consult with your physician and/or equipment provider and/or cannula manufacturer's instructions for
replacement information.
10.2 CASE CLEANING
WARNING
Liquid will damage the internal components of the concentrator and its equipment. To avoid damage
or injury from electrical shock:
• Remove the battery before cleaning
• Turn Off the concentrator and unplug the power cable before cleaning.
• DO NOT allow any cleaning agent to drip inside the air inlet and outlet openings.
• DO NOT spray or apply any cleaning agent directly to the cabinet.
• DO NOT hose down the product.
• DO NOT submerse the device or accessories in liquid
Harsh chemical agents can damage the concentrator and filters.
• DO NOT clean with alcohol and alcohol-based products (isopropyl alcohol), concentrated
chlorine-based products (ethylene chloride), and petroleum-based products or any other harsh
chemical agents.
• Mild liquid dish detergent is recommended.
Periodically clean the case as follows:
1. Make sure the concentrator is off, is removed from the carry bag, and the power cord or battery is
removed.
2. Clean the outside case using a cloth dampened with a mild liquid detergent and water.
3. Allow the concentrator to air dry, or use a dry towel, before returning the concentrator to the carry bag
or backpack and prior to operating the concentrator.
IMPORTANT: The device should receive an external cleaning weekly; accessories should be cleaned as
needed. The device is provided non-sterile and exterior should be cleaned and the output filter replaced
prior to delivering to a new patient.
©2024 Inogen, Inc. All rights reserved.
Page 27 of 120
96-13115-50-01 B

Publicité

loading