Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
EC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. main voltage prior to application of the test level.
12.5 GUIDANCE AND MANUFACTURER'S DECLARATION –
ELECTROMAGNETIC EMISSIONS
The concentrator is intended for use in home, institution, vehicle and other transportation and mobile envi-
ronments. The user of the concentrator should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
©2024 Inogen, Inc. All rights reserved.
IEC 60601 Test Level
3 Vrms
150 kHz to 80 MHz
6Vrms ISM and
amateur frequencies
10V/m
80 MHz
to 2.7 GHz
± 8 kV contact
± 2, 4, 6, 8 and
15 kV air
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
0% UT for 0.5 cycle at
0 °, 45 °, 90 °, 135 °,
180 °, 225 °, 270 °, and
315 °.
0% UT for 1 cycle
70% UT for 25/30 cycle
0% UT for 200/300 cycle
30 A/m
Compliance
Group 1
Electromagnetic Environment Guidance
The Rove 6 Portable Oxygen Concentrator is suitable
for the electromagnetic environment of typical home,
institution, vehicle, train, airplane, boat and other
transportation environments.
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should
be at least 30%.
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical home,
institution, vehicle or other transpiration and mobile
environments.
Mains power quality should be that of a typical
home, institution, vehicle and other transportation
and mobile environments. If the user of the Rove 6
requires continued operation during power mains
interruptions, it is recommended that the device be
powered from an uninterrupted power supply.
Power frequency magnetic fields should be at levels
characteristic of a typical home, institution, vehicle
and various mobile environments. Power frequency
magnetic fields from common appliances in the home
are not expected to affect
the device.
Electromagnetic Environment
Guidance
The concentrator uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and not likely to cause any
interference in nearby equipment.
Page 34 of 120
96-13115-50-01 B