Inogen Rove 6 Portable Oxygen Concentrator (Model # IO-501)
Operating atmospheric pressure
Operating altitude**
Shipping and storage temperature -13 to 158˚F (-25 to 70˚C)
Shipping and storage humidity
Measurement uncertainties:
Intelligent Delivery Technology
*Based on atmospheric pressure of 101.3 kPa (14.69 psi) at 20˚ C (68˚ F) & Dry (STPD).
**Operating outside of these operational specifications can limit the concentrator's ability to meeting
Oxygen Concentration specification at higher liter flow settings.
12.2 PULSE VOLUME FLOW SETTINGS
Inogen Rove 6 Pulse Volumes per Flow Setting
(mL/breath ± 15% per ISO 80601-2-67)
BREATHS PER MINUTE
10
15
20
25
30
35
40
TOTAL VOLUME PER MINUTE (ML/MIN) 210
CAUTION
• The setting of other models or brands of oxygen therapy equipment may not correspond with the
settings of this device.
• The settings of this device may not correspond with the setting for devices that provide continuous
flow oxygen.
©2024 Inogen, Inc. All rights reserved.
70 kPA to 106 kPA
0 to 10,000 ft (0 to 3048 meters)
Up to 90%, non-condensing
Store in a dry environment.
Pulse volumes: ± 15% of rated volume
Pressure: ± 0.03 psig (General) / ± 0.05 cm H2O (Inspiratory Trigger
Sensitivity)
Oxygen concentration: ± 3% (not accounting for temperature,
barometric pressure, and time from measurement device
calibration)
Inogen's devices use complex algorithms that are designed to
®
detect shallow breathing down to 0.12 cm H20 and will change the
bolus size of oxygen to meet the patient's breathing rate.
Upon detection, the Inogen delivers oxygen within the first 250
milliseconds of inspiration, when oxygen therapy is most effective.
1
21.0
14.0
10.5
8.4
7.0
6.0
5.25
Page 32 of 120
2
3
4
42.0
63.0
84.0
28.0
42.0
56.0
21.0
31.5
42.0
16.8
25.2
33.6
14.0
21.0
28.0
12.0
18.0
24.0
10.5
15.75
21.0
420
630
840
5
6
105.0
126.0
70.0
84.0
52.5
63.0
42.0
50.4
35.0
42.0
30.0
36.0
26.25
31.5
1050
1260
96-13115-50-01 B