Battery charging
When the Headphones battery is running low, the light beside the
Headphones power button will blink red and the Headphones will
emit a sequence of low-pitch beeps.
An empty battery should take no more than 3 hours to fully recharge, and
a full charge should allow the Headphones to be used for at least 8 hours.
Once the battery is fully charged, the Charger should be disconnected
to help conserve the life of the battery in the Headphones.
Chapter 11 - Device setup
To commence a treatment session, the Lenire device should be set up as follows:
11.1 - Controller and Headphones setup
The Headphones should be worn so they cover both ears and should be adjusted to maximise
comfort before commencing the treatment session.
It is essential that the left and right
earphones cover the left and right ears,
respectively. This is because the device
calibration is unique to each ear – it is
based on your audiogram. Failure to put on
the Headphones correctly could result in
excessive loudness in one ear, which could
cause discomfort, or incorrect loudness
that may reduce the effectiveness of the
treatment.
The left and right sides of the earphones
are marked L and R respectively on the
cloth inside the foam cushions.