Setting the station code (ID code)
If the optional USB port is being used, several WD 3M repair stations
can be activated and remotely controlled with their full range of
functions. For this purpose, each station requires a station code (ID
code) for unambiguous identification.
1. Select the menu item REMOTE ID in Menu 2.
2. Enter an ID with the UP or DOWN button
(possible values 0 – 999).
3. Proceed to the next menu item with the button ┌ 1 ┐ (back) or
┌ 2 ┐ (forward).
Note
Press button ┌ 3 ┐ to exit the menu item without changes (EXIT).
Using the calibration function (Factory Calibration Check = FCC)
With the FCC function you can check the temperature precision of
the repair station and compensate for possible deviations. For this
purpose, the soldering-tip temperature must be measured with an
external temperature meter and a temperature measuring tip
assigned to the soldering tool. The corresponding channel must be
selected prior to calibration.
Changing calibration at 100 °C / 212 °F
1. Insert the temperature sensor (0.5 mm) of the external
temperature meter into the temperature measuring tip.
2. Select the menu item FCC in Menu 2.
3. Press the DOWN button.Calibration point 100 °C / 212 °F is
selected.
The soldering tip is now heated to 100 °C / 212 °F.
The control indicator flashes as soon as the temperature is
constant.
4. Compare the displayed temperature of the measuring instrument
with that shown in the display.
5. Use the UP or DOWN button to set the difference between the
value indicated on the external meter and the value indicated on
the repair station.
Maximum possible temperature adjustment ± 40 °C (± 72 °F).
Example:
Display 100 °C, external measuring instrument 98 °C:
setting
Display 100 °C, external measuring instrument 102 °C:
setting
Note
Press button ┌ 3 ┐ to exit the menu item without changes (EXIT).
6. Press button ┌ 2 ┐ (Set) to confirm the value.
The temperature deviation is now reset to 0.
Calibration at 100 °C / 212 °F is now concluded.
7. Exit Menu 2 with button ┌ 3 ┐.