Use; Using The Products - Ugin MINICLEAN 2 Mode D'emploi

Polissage electromagnetique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Instructions for use
(rev 08/18)

5. USE

 Prepare the cleaning bowl: place inside the bowl the piece to be cleaned, the needles and the
water - cleaning product mixture (for example, a washing-up liquid)
 Replace the cap and cover and secure it tightly
 Position the closed cleaning bowl in the space provided for this purpose in the upper part of
MINICLEAN 2
 Activate the device through ON/OFF switch located in the back panel
 Select the cleaning time using the electronic timer (0 - 30 minutes) located on the front panel
(buttons + and -)
 Press START. The cycle starts up after a few seconds and stops when the timer has returned to
zero. To stop it before the end of the programed time, press STOP.
NOTE: In the course of the cleaning cycle, the direction of rotation of the magnetic field dis
reserved at regular intervals, each time after a short pause.
 Switch off the device through the ON / OFF switch
Take the piece cleaned into cleaning bowl. Use pliers and/or suitable gloves (follow the
cleaning product manufacturer's instructions: refer to the product data sheet and/or safety data
sheet)
 Rinse thoroughly in running water the pieces cleaned
 Rinse the bowl and empty out the water after each use
ATTENTION: after cleaning, the pieces must be rinsed thoroughly in running water.
Rinse the bowl and empty out the water after each use

5.1. USING THE PRODUCTS

The machine is delivered with:
- 2 pack 50g of needles ECOINOX
- 1 cleaning bowl in diameter 100mm (with cap and cover)
To ensure good results, it is important to use the proportions indicated below:
1) 50g of needles ECOINOX
2) about 5mL of cleaning product (for example, a washing-up liquid)
N.B.: quantity to be adjusted according to the product used
3) water to two-thirds of the bowl height
Before using the appliance for the first time, we recommend washing the ECOINOX needles in the
cleaning bowl containing a solution of water and about 5mL of cleaning product for 5 to 10
minutes.
N.B.: needles that have not been used for a prolonged period of time may darken in colour. In
such a case, simply prepare a product + water mixture and run a cleaning cycle for a few minutes.
After rinsing, the needles will look like new again.
UGIN DENTAIRE
MINICLEAN 2
13

Publicité

Chapitres

Table des Matières
loading

Table des Matières