Rev 290319 ARTIS NEO ARTIS NEO Déclaration de conformité CE – EC Declaration of conformity ....... 2 Description de l’appareil – Equipment description ........3 MODE D’EMPLOI ................... 6 INSTRUCTIONS FOR USE ................22 Visuels non contractuels / Non-contractual images...
Rev 290319 ARTIS NEO B. Description de l’appareil – Equipment description a. Description de la face arrière – Rear panel N° Description Description Interrupteur général M/A Main ON/OFF switch Prise alimentation secteur (230V) Mains power outlet (230V) Raccordement de vide...
Mode d’emploi (rev 290319) ARTIS NEO Déclaration de conformité CE – EC Declaration of conformity ....... 2 Description de l’appareil – Equipment description ........3 a. Description de la face arrière – Rear panel ............3 b. Description interface utilisateur – Description of programming unit ....4 MODE D’EMPLOI ...................
Ne pas effectuer d’opérations d’entretien différentes de celles indiquées dans ce manuel. Toute intervention non prévue dans la documentation peut être dangereuse. Pour toute information concernant l’installation, l’entretien et l’utilisation de l’appareil s’adresser au Service après-vente UGIN DENTAIRE. UGIN DENTAIRE...
- Choix des affichages des paramètres en 5 langues (français, allemand, espagnol, anglais, italien) 3. DESEMBALLAGE Vérifier au déballage le parfait état du four ARTIS NEO et de l’ensemble des accessoires Signaler au livreur d’éventuelles détériorations Ne pas oublier de faire signer par le livreur le bon de transport...
Mode d’emploi (rev 290319) ARTIS NEO 4. INSTALLATION ET MISE EN SERVICE 4.1. Installation Placer le four dans un endroit aéré en respectant un espace minimum de 10 cm avec toute autre paroi. IMPORTANT: il est recommandé de faire fonctionner le four sur un circuit électrique indépendant...
Mode d’emploi (rev 290319) ARTIS NEO 5. PROGRAMMATION 5.1. Ecran alphanumérique L’écran peut présenter deux états, bien distincts: A. En programmation (obtenu par la touche PROGRAMATION Il permet d’introduire ou de modifier les paramètres d’un programme et se caractérise par le clignotement d’un des paramètres.
Page 12
Mode d’emploi (rev 290319) ARTIS NEO 5.3. Touche STAR/STOP Cette touche permet de lancer le cycle de cuisson. Une deuxième action en cours de cycle interrompt la cuisson et le four revient en position « ATTENTE » (plateau en position basse).
Mode d’emploi (rev 290319) ARTIS NEO Page 3 Page 4 Page 5 5.5. Fleche vers la droite - Pendant la programmation, cette touche permet d’accéder au paramètre suivant - Pendant le cycle de cuisson, elle annule la phase en cours et permet de passer à la suivante 5.6.
Mode d’emploi (rev 290319) ARTIS NEO 5.7. Fleche vers le bas - En position « ATTENTE » : change le n° de programme - Lors de la programmation : diminue la valeur du paramètre qui clignote - En cours de cuisson : annule le cycle pour revenir en position « ATTENTE » (plateau en position basse) 5.8.
Niveau de vide : 0 à 99% La carte de commande du four ARTIS NEO est équipée d’un capteur de vide électronique. Plus précis et plus fiable qu’un manomètre traditionnel, il permet de réguler le niveau du vide dans un créneau de ±...
Page 16
Mode d’emploi (rev 290319) ARTIS NEO Rampe : 0 à 200°C/min L’élévation en température est parfaitement linéaire du début jusqu’à la fin de la cuisson. Sa valeur est réglable de degré en degré. La vitesse de montée est respectée, même si l’alimentation électrique varie dans une limite de ±...
Mode d’emploi (rev 290319) ARTIS NEO 7. PARAMETRES SECONDAIRES Cette deuxième série de paramètres peut être modifiée. Pour y accéder lorsque le four est en position « VEILLE » ou « ATTENTE », appuyer simultanément sur les touches . Seule une partie de ces paramètres concerne les données techniques de l’appareil en rapport avec les fonctions de cuisson.
Mode d’emploi (rev 290319) ARTIS NEO 8. CALIBRATION TEMPERATURE Le four a fait l’objet d’un contrôle de température à l’aide d’un calibrateur de température. Lors de l’étalonnage en usine et lorsque le moufle est neuf, le réglage est exacte (indice = 5).
ARTIS NEO 9. ENTRETIEN ET PRECAUTIONS D’UTILISATION Lorsque des travaux de nettoyage et/ou de maintenance sont effectués, le four ARTIS NEO doit être déconnecté de l'alimentation électrique principale, et à température ambiante : laisser refroidir après utilisation (risque des brûlures et/ou dommages matériels) Hormis les nettoyages extérieurs, ARTIS NEO se limite aux opérations suivantes :...
Refaire un cycle « à vide » pour vérification. ATTENTION : la liste ci-dessus est partielle, certains affichages peuvent combiner plusieurs défauts. Exemple : PANNE 22 PANNE 16+4+2 Dans tous les cas, il est impératif de CONTACTER LE S.A.V. UGIN DENTAIRE UGIN DENTAIRE...
SURTENSIONS L’alimentation d’un four à microprocesseur nécessite un filtrage parfait. Le four ARTIS NEO est protégé contre les surtensions par un système de filtres actifs et passifs. Cependant, dans le cas de surtensions élevées, il peut y avoir un risque de perturbation des programmes.
Mode d’emploi (rev 290319) ARTIS NEO 13. TABLEAU DE CONVERSION DES UNITES DE VIDE % de vide mm Hg in Hg mbar 1.49 50.663 2.99 101.33 4.48 151.99 5.98 202.65 7.48 253.31 8.97 303.98 10.47 354.64 11.96 405.3 13.46 455.96 14.96...
Page 23
Instructions for use (rev 290319) ARTIS NEO Déclaration de conformité CE – EC Declaration of conformity ....... 2 Description de l’appareil – Equipment description ........3 a. Description de la face arrière – Rear panel ............3 b. Description interface utilisateur – Description of programming unit ....4 INSTRUCTIONS FOR USE ................
Do not execute different operations of maintenance from those brought back in the manual. Whichever operation not included in this manual, can involve risks. For whichever inherent information concerning installation, maintenance and use of the unit, contact the customer’s service UGIN DENTAIRE. UGIN DENTAIRE...
- Choice of five languages for parameter displays (French, German, Spanish, English or Italian) 3. UNPACK Unpack ARTIS NEO and check that it is in perfect order, so that any defects can be reported to the carrier. Remember to get the carrier to sign the delivery note...
Instructions for use (rev 290319) ARTIS NEO 4. INSTALLATION AND START-UP 4.1. Installation Install the furnace in a well-ventilated area and make sure it is at least 10 cm from walls or other surfaces. IMPORTANT: We recommend that the furnace be operated on a separate electric circuit.
Instructions for use (rev 290319) ARTIS NEO 5. PROGRAMMING 5.1. Alphanumeric display There are two different display modes, which must be clearly differentiated: A. The programming mode (initiated by pressing the key PROGRAMING In this mode, you can select or modify the parameters in a program, which are displayed flashing.
Page 28
Instructions for use (rev 290319) ARTIS NEO 5.4. PAGE key When programming (push on ), it allows to pen the 5 pages of parameters Page 1 It can be modified by pushing the keys , that gives access to the alphabet (capital and minuscule) as well as number from 0 to 9.
Page 29
Instructions for use (rev 290319) ARTIS NEO Page 4 Page 5 5.5. Arrow key to right - During programming, press this key to switch from one parameter to the next - During a firing cycle, press this key to cancel the ongoing phase and move on to the next 5.6.
Page 30
Instructions for use (rev 290319) ARTIS NEO 5.9. Programming key Use this key to initiate the programming mode. After pressing , you can enter or modify the firing parameters. Once the program is entered, press again to "confirm" the program settings.
Vacuum level: 0 to 99% The programming unit on the ARTIS NEO furnace has an electronic vacuum sensor. As it is more accurate and reliable than a conventional pressure gage, the vacuum level can be adjusted to within ± 3% of the programmed value. Unit conversion: see chapter 13.
Page 32
Instructions for use (rev 290319) ARTIS NEO Heat-rise rate: 0 to 200°C/min The temperature rises at a perfectly linear rate from the beginning to the end of the firing cycle. You can adjust the temperature degree by degree. The heat-rise rate is maintained, even if the supply voltage varies within a maximum of ±...
Instructions for use (rev 290319) ARTIS NEO 7. SECONDARY PARAMETERS This is a second series of parameters that can be modified. These parameters correspond to the appliance's technical data, but only some of them affect the firing functions. To access the latter when the furnace is on "STAND-BY" or “IDLE” position: press both the keys at the same time.
Instructions for use (rev 290319) ARTIS NEO 8. TEMPERATURE CALIBRATION The furnace have been subjected to a temperature test using a temperature calibrator. The setting is exact (SET VALUE = 5) during factory calibration and while the muffle is new. However, should you not find the initial setting satisfactory;...
(risk of burns and/or damage) Except for cleaning the machine outside, ARTIS NEO furnace maintenance entails only two other operations: - Cleaning the platform sealing joint regularly with alcohol...
Instructions for use (rev 290319) ARTIS NEO 10. TROUBLE-SHOOTING Type of fault displayed continuous alarm sounds: operator action required First stop the alarm by pressing the key Depending on the type of fault displayed, consult the list below to determine the cause or causes of failure and contact your After-Sales Service ...
Instructions for use (rev 290319) ARTIS NEO 11. PROTECTION PROGRAMS Your programs are protected by an EEPROM memory, which operates without batteries. OVERVOLTAGE The power supply to furnaces with micro-processing units requires perfect filtering. The ARTIS NEO furnace is protected against over-voltage by a system of active and passive filters. However, very high over-voltages may interfere with your programs.
Instructions for use (rev 290319) ARTIS NEO 13. VACUUM UNITS CONVERSION TABLE % vacuum mm Hg in Hg mbar 1.49 50.663 2.99 101.33 4.48 151.99 5.98 202.65 7.48 253.31 8.97 303.98 10.47 354.64 11.96 405.3 13.46 455.96 14.96 506.63 16.45 557.29...