en
|
Troubleshooting
Byte
Bit
Description
21
1
SIMULATION_ACTIVE
2
SOFTWARE_UPDATE_INCOMPLETE
3
POWER_LOW
4
POWER_CRITICAL
5
FAULT_PREDICTION
6
ENVIRONMENTAL
7
OUTSIDE_LIMITS
22
0
REAL_TIME_CLOCK_NOT_VALID
1
WARNING_UCR_VOLTAGE
23
6
OTHER_FAULTS
24
3
NO_NETWORK_CONNECTION
4
SENSOR_VITALITY_POOR
5
SENSOR_CAL_INTERVAL_EXPIRED
6
SENSOR_WARMUP
54
Instructions for use
|
Dräger Polytron
Cause
Device is in simulation mode.
Firmware update active.
Firmware update failed.
Battery charge low.
Battery charge critical.
Device warning.
Device is operated outside the specified tempera-
ture range.
Measured value too low.
®
No connection to the WirelessHART
network.
Error in sensor.
Serious device fault, various causes.
®
Incorrectly configured WirelessHART
network.
Vitality of the sensor is low.
Calibration interval expired.
EC sensor in warm-up phase 1.
®
6100 EC WL, Dräger Polytron
Remedy
Exit simulation mode via PolySoft.
Wait for firmware update (maximum 15 minutes).
A
Restart the device.
B
Have the gas detector checked by DrägerService.
Change battery pack soon.
Change battery pack immediately.
®
A
Check WirelessHART
connection.
B
Check detail status with PolySoft.
In case of high temperatures due to solar radiation,
use weather protection set (accessory).
A
Calibrate the sensor.
B
Change the sensor.
®
Connect to WirelessHART
network.
Change the sensor.
Have the gas detector checked by DrägerService.
®
Check WirelessHART
network settings.
Change sensor soon.
Calibrate the sensor.
Wait for warm-up phase 1 to finish.
®
Repeater WirelessHART
®